This
section presents the paper currently in preparation. The authors are :
-
Philippe Giaccari, Omar Sqalli and
Hans G. Limberger : Institute of Applied Optics, Swiss Federal Institute
of Technology
Scanning near-field optical
microscopy requires a performant sensor to measure the tip-to-sample distance.
In this letter, we report on a novel shear force detection scheme for scanning
near-field optical microscopy applications. It is based on an all fiber
low-coherence interferometer. This setup makes possible the measurements of the
tip oscillation amplitude of less than 50 pm both in air and aqueous
environment with a precision of 160 fm. Hz-1, thus demonstrating the
ability to perform topographic measurements both in air and in liquids with a
resolution better than 1 nm in the height direction. Stable feedback in air and
fluids is obtained with tip-sample interaction forces below 1 pN.
Scanning
near-field optical microscopy (SNOM) has drawn considerable research interest
in recent years since it allows the measurement of both the topography and the
optical contrast of a sample with sub-wavelength resolution [7-1]. The
instrument works by scanning a sub-wavelength size probe very close to the
sample surface. The probe consists of a glass tip that can be covered with an
opaque metal layer, with a clear aperture of sub-wavelength dimension at the
tip apex [7-2]. A large majority of today's probe-sample distance control mechanisms
works by detecting the damping of the oscillating tip by lateral forces (the so
called "shear force") close to the surface [7-3]. Scanning-near-field
optical microscopes require consequently performant sensors to measure
nanometric oscillations of SNOM tips. Different methods have been proposed in
the last ten years, such as a compact near-field optical module based on an
external cavity laser interferometer [7-4] or the tuning fork [7-5,7-6] that
allows measuring tip oscillation amplitudes of a few picometers in air.
However, working in aqueous environment is more critical since a huge damping
of the tip vibrations occurs upon immersion of several microns into liquid
[7-7]. For investigations in aqueous environment, a new type of liquid cell was
proposed, in order to limit the immersion depth of the SNOM probe [7-8].
Measurements in aqueous environment with the tuning fork system are possible
only upon complete immersion of the tuning fork into liquid [7-9] with a
typical vibrating tip amplitude of about 1 nm.
In
this letter, a novel force detection scheme for scanning near-field optical
microscopy applications is presented, theoretically described and
experimentally tested. It is based on an all fiber low-coherence interferometer
to measure extremely small SNOM tips oscillation amplitudes, both in air and
aqueous environments.
The
novel force detection scheme for scanning near-field optical microscopy
applications is based on an all fiber low-coherence interferometer. The
experimental set-up is presented in Fig. 7-1. A super luminescent laser diode
(SLD) beam coupled into a single mode fiber (port 1) is used to illuminate a
Michelson interferometer based on a 50/50% fiber coupler. At the end of the
second interferometer arm (port 2), the measurement fiber, 4% of the light is
reflected at the glass-air interface. 96% of the light is transmitted and
partially reflected on the SNOM tip. The distance d between the SNOM tip and the end interface of the control fiber
is typically 2 microns. Therefore, the tip interface and the fiber end face
form a Fabry-Perot interferometer. A large part of the light reflected in this
structure is coupled back into the optical fiber and is detected with a
balanced detection system consisting of to photodiodes mounted at the other
arms (port 3 and 4) of the Michelson interferometer. Monitoring the intensity
of the interference fringes allows measuring the tip vibration amplitude. The
balanced detection scheme improves the S/N ratio by reducing the source noise.
The SLD source (SLD 56-MP SUPERLUM, 0.5 mW) spectrum has a full width at half
maximum Dl of 44 nm centered at lo=1319nm,
leading to a coherence length of the source of about 20 microns. The low
coherence of the SLD source has ths advantage to eliminate spurious interference
signals resulting from other reflections in the set-up (e.g., the coupler),
thus leading to an increase of the signal-to-noise ratio of 30 dB. The SNOM-tip
is mechanically excited by a piezoelectric element P2 located at x=0. The
excitation is being supplied by a digital Lock-In Amplifier (SRS, RF Lock-In
Amplifier, Model SR844). The measured optical interference signal is amplified
by the Lock-in Amplifier and finally sent to a PC for storage and display.
Fig. 7-1. Schematic drawing of the experimental
set-up of the shear-force system based on low coherence interferometry. R1
(=4%) and R2 (=96%) are the reflection coefficients at the end of the control
fiber and the SNOM tip, C a 50/50 optical coupler, D1 and D2 two detectors, p2
a dithering piezo.
In
order to calculate and characterize the SNOM fiber tip oscillations, we
consider the vibration model of a beam clamped at one end and free at other
[7-10]. The tip is described as a homogenous quartz cylinder, since the 100
microns long conical part of the tip is insignificant in comparison to several
millimeters long cylindrical fiber. We thus consider a uniform radius R
of 62.5 mm along the entire SNOM fiber length. The mass per unit length of the
quartz is 8.6 mg/m, E the Young modulus is 72 GPa. For a given harmonic
excitation frequency and a given fiber length, the vibrations amplitude at a
distance from the clamped end is calculated by resolving the Euler fourth order
differential equation [7-10]. Fig. 7-2 shows (a) the measured and (b) the
calculated oscillations amplitudes at the middle of a 9.2 mm long quartz fiber
that has a diameter of 125 mm. We observe six resonances with different
amplitudes that correspond to the vibration modes. The precision of the
vibration amplitude measurement is 160 fm/Hz-1/2. The calculated
resonance positions and relative amplitudes are similar to the experimental
measurements.
Fig.
7-2. Measured (a) and calculated (b) vibration amplitudes at the middle
(x=4.6mm) of a 9.2 mm long SNOM tip.vibration amplitudes at the middle
(x=4.6mm) of a 9.2 mm long SNOM tip.
Fig.
7-3. Calculated vibration amplitudes of a 9.2 mm long SNOM tip as a function of
the position x on the tip and the oscillation amplitude.
Fig.
7-3 illustrates the tip oscillations amplitudes calculated at a position x on
the tip and as a function of the oscillation frequency. We observe six
resonances with different amplitudes that correspond to the vibration modes.
The theoretical calculations allow correctly estimating the oscillations at the
end of the tip by measuring the oscillation amplitude in another part of the
fiber tip, for a given eigenmode. The above described set-up makes possible to
detect a minimal vibration amplitude of the SNOM tip of about 5 pm for a
lock-in time constant of 1 ms, and of 1 pm for a time constant of 30 ms.
Near-field optical microcopy measurements are consequently performed with
typical oscillation amplitudes of 50-100 pm at the tip extremity, and a signal
to noise ratio always superior to 10.
The
vibration modes of the tip are experimentally investigated in aqueous solution.
Fig. 7-4 illustrates the vibration amplitude measurements of 6.2 mm long quartz
SNOM tip in air and in water, for different immersion depths of the tip in
water. The vibrations measurements have been carried out at the middle of the
fiber tip for the third eigenfrequency. First, a damping of the vibrations
amplitude as well as a shift of the resonance frequency to lower values is
observed when the immersion depth increases. The resonance position is shifted
from 46.8 kHz to 44.5 kHz for a 2 mm immersion depth. Second, the Q factor
decreases from 100 to 80 but remains always sufficiently high and makes
possible performing SNOM measurements in water, even for immersion depth of 2
mm. The same behavior is observed for the other resonances showing that the
vibration modes are preserved in water.
Fig.
7-4. Damping of the L=6.2 mm long tip
oscillations, as a function of the oscillation frequency, for several tip
immersion depths in water. The measurement is performed at x=3.1 mm, at the
middle of the oscillating SNOM tip.
Fig.
7-5. Two topographic images of a 21 nm deep chromium on glass grating with a
period of 372 nm performed with the same
tip on the same sample in air then in water.
The
previously described interferometric system is mounted in the SNOM set-up. A
z-piezo vertically moves the tip, whereas an x-y piezo horizontally moves the
sample. Fig. 7-5 shows two topographic slices of chromium on glass grating with
a period of 372 nm performed with the same tip, in air and in water. The tip
vibration amplitude at the tip extremity of 50 pm during the scan in both
cases, with a signal-to-noise ratio was of about 50. The similarity of the
images proves the reliability of the technique to perform accurate measurements
both in air and aqueous environment with a height precision better than 1 nm.
Moreover, the topographical contrast is nearly the same in air and in water.
The lateral resolution is given by the probe shape.
To
gain a better qualitative and quantitative understanding of the interaction
force between the tip and the sample, a simple model called the effective mass
harmonic model and described in reference [7-5,7-8] is used. Again, the SNOM
fiber tip is considered as an uniform cylinder with a static spring constant kspring.
The tip-sample interaction shear-force Fint is obtained by measuring the free Ul
and attenuated Uint vibration amplitude, at a specific resonance
frequency with a precise Q quality factor:
. The
SNOM tip described in Fig. 7-3 has a length L of 9.2 mm, a spring
constant kspring=3EI/L3 of 3 N/m (fundamental
eigenfrequency), where I=pR4/4 is the inertia mement of the tip, a working free
oscillation amplitude Ul at 2 kHz of about 50 pm, a Q factor of
about 80. By choosing Uint equal to 0.9*Ul,
the measured shear force Fint is about 0.2 pN. Note that an increase
of the probe length leads to a decrease of the probe static spring constant,
and therefore to the detection of a smaller shear force for the same vibration
amplitude. Higher eigenfrequencies are usually characterized by a higher Q
factor, that allows measuring smaller forces, but also a higher spring constant
of the tip, since the nodes reduce the effective oscillating length of the tip.
In conclusion, a new low coherent system has
been implemented in force detection schemes for scanning near-field optical
microscopy applications. It allows characterizing the SNOM-tip oscillation
modes and amplitudes on the one hand, and, on the other hand, performing
topographical measurements with a high precision both in dry and aqueous
environments using the shear-force technique. The SNOM tip vibration amplitudes
are typically 50-100 pm at the tip extremity during the scan. Topography
measurements with a precision better than 1 nm in the z direction were
performed without any control of the ambient temperature and humidity.
|